• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO XL; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO XL; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00547120
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: date of event was approximated to (b)(6) 2018 as no event date was reported.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was unpacked for a procedure on an unknown date.According to the complainant, during the preparation, it was noted that a hair was found inside the package.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was unpacked for a procedure on an unknown date.According to the complainant, during the preparation, it was noted that a hair was found inside the package.
 
Additional Manufacturer Narrative
Date of event: date of event was approximated to 10/01/2018 as no event date was reported.Problem code (b)(4) captures the reportable event of foreign matter.Investigation result a visual examination of the complaint device revealed that the device did not have visual defects.There was no hair observed in the pouch and the pouch was opened.The inner radio opaque (ro) markers inside the balloon were inspected for any sharp edge but none of them showed abnormalities.Based on the analysis performed and the information provided, the most probable root cause for the complaint event is no problem detected since the device complaint or problem cannot be confirmed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO XL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8029296
Report Number3005099803-2018-61166
Device Sequence Number1578753
Product Code FGE
UDI-Device Identifier08714729790648
UDI-Public08714729790648
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2020
Device Model NumberM00547120
Device Catalogue Number4712
Device Lot Number0022270132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2019
Initial Date Received by Manufacturer 10/08/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date02/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-