Model Number MMT-723NAP
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/23/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had keypad anomaly, display anomaly and motor error.The customer¿s blood glucose level was unknown.The customer reported that the buttons were not working properly.The customer reported that they received a motor error alarm.Customer was able to successfully clear the alarm.Customer was able to complete pump rewind.The customer reported that the drive support cap was normal.It was reported that the screen was scratched and forced to angle pump to see display and reservoir compartment opening was chipped.The customer reported that the time clock was advancing.The customer was advised to discontinue use of the pump and revert to backup plan per health care professional instructions.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
All buttons functioning properly.No moisture/damage/unlocked lcd keypad connector during visual inspection noted.Device display properly.No missing segment/partial display anomaly noted during testing.No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.Device had minor scratched lcd window, cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8030036
Report Number2032227-2018-54204
Device Sequence Number13751763
Product Code OYC
UDI-Device Identifier00613994904348
UDI-Public(01)00613994904348
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAP
Device Catalogue NumberMMT-723NAP
Device Lot NumberA5723NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2019
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer03/29/2019
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date04/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient Weight118
-
-