| Model Number |
MMT-723NAP |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/23/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had keypad anomaly, display anomaly and motor error.The customer¿s blood glucose level was unknown.The customer reported that the buttons were not working properly.The customer reported that they received a motor error alarm.Customer was able to successfully clear the alarm.Customer was able to complete pump rewind.The customer reported that the drive support cap was normal.It was reported that the screen was scratched and forced to angle pump to see display and reservoir compartment opening was chipped.The customer reported that the time clock was advancing.The customer was advised to discontinue use of the pump and revert to backup plan per health care professional instructions.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
All buttons functioning properly.No moisture/damage/unlocked lcd keypad connector during visual inspection noted.Device display properly.No missing segment/partial display anomaly noted during testing.No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.Device had minor scratched lcd window, cracked reservoir tube lip.
|
| |
|
Search Alerts/Recalls
|