MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550R
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that the endoscope was blurry.It is unknown when this event occurred or if the surgery was successful.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Event or Problem Description
New information was received indicating that the event occurred during setup.No patient effect.Product was change out.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 1, 2018.(b)(4).To check the performance of the endoscope, a visual observation of the internal components using a monocular was conducted.It was discovered that there was a crack on one of the internal lenses which resulted in a blurry visual field.Possible cause might be the product was either dropped or inadvertently struck by an object or an the excessive force was exerted on the endoscope shaft; this could deflect the shaft and lead to the alteration of the internal lens system.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED ENDOSCOPE 5.5M
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8030760
Report Number1124841-2018-00277
Device Sequence Number1499110
Product Code GCJ
UDI-Device Identifier00699753023222
UDI-Public(01)00699753023222
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160206
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMCENDO550R
Device Catalogue NumberN/A
Device Lot Number740014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/08/2018
Supplement Date Received by Manufacturer11/06/2018
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-