Model Number 7040-90
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Arthralgia (2355)
Date of Event 10/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper implant size selection, using an implant that was too short.
 
Event or Problem Description
The patient had left side si joint arthrodesis in (b)(6) 2017 where three implants were placed.The patient later complained of little pain relief.The surgeon thought that the middle implant may have been too short.In (b)(6) 2018, the surgeon performed a revision surgery where he removed the middle implant using chisels as it was solidly fixed in bone and replaced it with a larger implant pf the same type.He then added an additional implant, again of the same type, and bone graft to the joint to help fixate the joint.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key8030786
Report Number3007700286-2018-00097
Device Sequence Number11109364
Product Code OUR
UDI-Device Identifier00859256003286
UDI-Public00859256003286
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2021
Device Model Number7040-90
Device Lot Number493739
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2018
Initial Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
-
-