| Model Number |
AR40E |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that when being inserted into the eye an ar40e intraocular lens (iol) had broken haptic.It was stated that the trailing haptic was pinched and nearly broken when implanted.Iol was removed from the patient and another lens was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 10/18/2018.Device returned to manufacturer ¿ yes.Device evaluation: the device was received at the manufacturing site for investigation.Only half of the lens was received with the haptic detached.Residue of viscoelastic was detected in the half of the lens received.The condition observed is consistent with a lens that has been explanted.The reported issue was not verified however, a haptic detached was confirmed but it could not be determined if the condition observed is related to manufacturing process as the reported device was handling in a surgical procedure.Based on the returned product evaluation, a product quality deficiency or product malfunction could not be determined.Manufacturing record review: the manufacturing record review of the production order was evaluated, and the devices were manufactured within specifications.The units were released according to specification.Historical data analysis: a search of complaints was performed.The search revealed that no other complaint has been received for this production order number related to this investigation.Labeling review: a review of the directions for use (dfu) and precautions was conducted.The dfu instructs the customer to inspect the product before use for any damage.The information in the document is clear and no request for potential update is required.Conclusion: as a result of the investigation a product quality deficiency or product malfunction could not be determined.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|