• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Failure to Pump (1502)
Patient Problem No Patient Involvement (2645)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility for annual maintenance. During the service activity the reported issue was identified. The technician traced the failure to a defective patient pump. However, the technician could not determine how the defect was caused. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a patient pump was found to be not working. This issue was identified by a livanova field service representative during routine maintenance. There was no patient involvement.
 
Manufacturer Narrative
The failing part was requested to livanova deutschland for investigation. Results revealed that the pump was dirty and that the pump motor board was defective. The most likely root cause of the damage is water ingress since lime residues are visible on the pump motor.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8031843
MDR Text Key125923268
Report Number9611109-2018-01314
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/12/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-80
Was Device Available for Evaluation? No Answer Provided
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/15/2019
Is This a Reprocessed and Reused Single-Use Device? No

-
-