| Model Number |
48-50-00 |
| Medical Device Problem Code |
Loss of Power (1475)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative tested the s5 system and no failure were identified; the device worked as intended.However, it was found that the battery of the device was deep discharged once and that the battery life was at 0%.A serial readout evaluation was performed but no specific issue could be identified.Therefore, livanova deutschland requested the e/p pack back for further investigation.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that a s5 system stopped working during procedure.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The e/p pack was returned to livanova (b)(4) for further investigation.During the evaluation the reported issue could be confirmed and was traced back to improper user handling as the device was not charged in a proper way.Therefore the system was running on battery mode with discharged batteries which led to the reported malfunction.
|
| |
|