| Model Number |
MMT-7020A |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer blood glucose was 162mg/dl and 128mg/dl and the sensor glucose was 61 mg/dl at the time of the incident.Insulin delivery was suspended due to sensor glucose values.Sensor glucose value that triggered the suspend event 61mg/dl.Threshold suspend limit in sensor settings 65mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The device will not be returned for analysis.
|
| |
|