• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1715KM 630G 3ML BLACK MEDI  ; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1715KM 630G 3ML BLACK MEDI  ; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KM
Device Problems Device Alarm System (1012); Loss of Power (1475); Failure to Power Up (1476); Failure to Read Input Signal (1581)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had keypad anomaly and replace battery alert.The customer¿s blood glucose level was 209 mg/dl.The customer reported that the replace battery alert with no low battery alert.The customer reported that the second occurrence of replace battery alert without a low battery warning.It was reported that they had unexpected battery power loss and inserted a new battery and now pump will not turn on.The customer reported that they were pressing the buttons and nothing was happened.The customer reported that they changed the battery and buttons were responding.The customer was advised that they may continue to wear pump until replacement arrives if they insert a new battery.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Insulin pump received with all buttons responding properly.No damage or moisture damage on keypad assembly.No unlocked keypad connector.No button keypad anomalies noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1715KM 630G 3ML BLACK MEDI  
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key8031898
MDR Text Key126004882
Report Number2032227-2018-54722
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169873834
UDI-Public(01)00643169873834
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715KM
Device Catalogue NumberMMT-1715KM
Device Lot NumberHG27PHG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age77 YR
Patient Weight59
-
-