Model Number MMT-1715KM |
Device Problems
Device Alarm System (1012); Loss of Power (1475); Failure to Power Up (1476); Failure to Read Input Signal (1581)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/25/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump had keypad anomaly and replace battery alert.The customer¿s blood glucose level was 209 mg/dl.The customer reported that the replace battery alert with no low battery alert.The customer reported that the second occurrence of replace battery alert without a low battery warning.It was reported that they had unexpected battery power loss and inserted a new battery and now pump will not turn on.The customer reported that they were pressing the buttons and nothing was happened.The customer reported that they changed the battery and buttons were responding.The customer was advised that they may continue to wear pump until replacement arrives if they insert a new battery.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Insulin pump received with all buttons responding properly.No damage or moisture damage on keypad assembly.No unlocked keypad connector.No button keypad anomalies noted.
|
|