• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).The intraocular lens was not implanted.If explanted; give date: n/a (not applicable).The intraocular lens was not implanted.(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the haptic of the pcb00 tecnis monofocal iol (intraocular lens) was observed broken upon insertion into the patient's eye.There was patient contact, but the surgeon was able to retrieve the lens before fully implanting it.Another lens was successfully implanted and no further intervention was required.The patient was reported healing well.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 11/05/2018.Device returned to manufacturer? yes.Device evaluation: the preloaded delivery system, model pcb00, was returned at the manufacturing site for evaluation.The plunger was observed in partially advanced position.The pcb00 device was observed under the microscope and scarce amount of viscoelastic was observed on the cartridge.Dfu (direction for use) states to completely fill the viewing window of the pcb00 with ovd (ophthalmic viscosurgical device).The intraocular lens (iol) was observed stuck at the cartridge tip.There were no damages observed on the assembly.The customer's reported complaint was not verified, but, stuck in cartridge was confirmed.However, it could not be determined if the returned product condition was related to the manufacturing process as the device was handled and prepared for surgical procedure.Based on the returned product evaluation, product quality deficiency or product malfunction could not be determined.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8032002
Report Number2648035-2018-01424
Device Sequence Number19198357
Product Code HQL
UDI-Device Identifier05050474558434
UDI-Public(01)05050474558434(17)191027
Combination Product (Y/N)Y
Initial Reporter CountryGB
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/27/2019
Device Model NumberPCB00
Device Catalogue NumberPCB0000285
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer11/14/2018
10/25/2020
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date11/19/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-