MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 1884006EM
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The inner assembly turned by hand however the middle assembly tip rotated with it.The middle assembly cutting tip was detached at the spiral wrap, it was only resting on the inner cutter tip.The portion that broke off measured 0.45¿ long.There was cutting tooth damage on the inner and middle assemblies.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported via a manufacturer representative that the tip of the blade was broken into pieces during a functional endoscopic sinus surgery procedure.A backup device was not used because the procedure was almost completed however, there was a 30-minute delay due to this.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® SYSTEM NAVIGATED BLADE
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8032467
Report Number1045254-2018-00554
Device Sequence Number11128797
Product Code HAW
UDI-Device Identifier00613994617408
UDI-Public00613994617408
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K130608
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/04/2020
Device Model Number1884006EM
Device Catalogue Number1884006EM
Device Lot Number0215671438
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2018
Initial Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-