DAVIS & GECK CARIBE LTD SURGIPRO II; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Model Number VP-728-X
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 10/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a coronary artery bypass, after the anastomosis, while doing a continuous stitch on descending aorta, the suture broke at the strand (above the knot) and caused massive bleeding.The surgical time was extended by 30 minutes or more due to the problem.They immediately dressed the wound.They used same device from the same lot to complete the case.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of fifty eight representative samples of devices.Visual inspection and functional tes ting confirmed there were no abnormalities that would have caused or contributed to the reported condition.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIPRO II
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key8032552
Report Number9612501-2018-01913
Device Sequence Number1499241
Product Code GAW
UDI-Device Identifier10884521038714
UDI-Public10884521038714
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K050947
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberVP-728-X
Device Catalogue NumberVP-728-X
Device Lot NumberD8C0435SX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 10/08/2018
Supplement Date Received by Manufacturer12/18/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
-
-