| Model Number |
9-ITV10F45/80 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2018, a 24mm amplatzer septal occluder (lot number: 6184089) was deployed using a 10f amplatzer torqvue delivery system (lot number: 5790870).While repositioning the device, the la disc failed to be re-captured.The physician removed the device through the femoral vein and the case was aborted.The patient was taken to surgery to close the asd.The patient is reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of failure to recapture the left atrial disc could not be confirmed.The results of this investigation confirmed the device all functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
|
| |
|