• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 6X12 MILAGRO INTERFRENCE SCREW; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 6X12 MILAGRO INTERFRENCE SCREW; ORTHOPAEDIC BONE SCREW, BIOABSORBABLE Back to Search Results
Catalog Number 231813
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4)-incomplete the lot number is currently unavailable; therefore, the exp date is unavailable.The complaint device is not being returned as it was discarded by the customer; therefore, this complaint cannot be confirmed and duplicated.The assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the sales rep via phone that during a fhl transfer procedure, it was observed that milagro interference screw 6x12mm device broke mid-body while being inserted.The sales rep stated that the hole had been drilled and the screw broke while being tapped in.The screw was removed with no debris left behind.The case was completed with a new screw in the same bone hole with no patient harm or delay.The device was discarded by the customer.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Further, no lot numbers were supplied which precludes conducting a device history record (dhr) review or a lot specific search in the complaints handling system.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, (b)(4) will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).Depuy synthes has been informed that the lot number is not available.
 
Event or Problem Description
Additional information received from sales representative on (b)(6) 2018: the lot number was proved as 3914932; however, the consignment team indicated the lot provided was invalid.There were no procedural or patient anatomy factors which may have contributed to the breakage.The patient¿s bone quality was noted as normal.The insertion was not off axis.The same bone hole was used to complete the procedure.The tunnel and graft size were six millimeters (6mm).No trialing was performed.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A non-conformance search was performed for this part (231813) and lot (3914532) combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6X12 MILAGRO INTERFRENCE SCREW
Common Device Name
ORTHOPAEDIC BONE SCREW, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8033071
Report Number1221934-2018-55164
Device Sequence Number1643204
Product Code GEI
UDI-Device Identifier10886705021956
UDI-Public10886705021956
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K143660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number231813
Device Lot Number3914532
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer10/31/2018
11/07/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date11/07/2018
11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-