Model Number EEATAID25D
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, on creating the gastrojejunostomy during laparoscopic roux-en-y gastric bypass, a component of the device stayed with the patient.It was only determined after the operation and they had to be opened again.All pieces were retrieved and a computed tomography (ct) abdomen was needed to locate the components.They did a re-intervention to complete the case and there was an extension of inpatient stay for 2 days.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of a device.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8033098
Report Number1219930-2018-05800
Device Sequence Number1302690
Product Code FZX
UDI-Device Identifier20884521045719
UDI-Public20884521045719
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberEEATAID25D
Device Catalogue NumberEEATAID25D
Device Lot NumberK7J0003X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer01/14/2019
02/05/2019
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date02/04/2019
02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age27 YR
Patient Weight103
-
-