| Model Number |
6924 |
| Medical Device Problem Code |
Unintended Electrical Shock (4018)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is a late mdr report, due to late reporting of event information from our field office.The reported device will be investigated upon its return to microline surgical, inc., to submit a 30-day follow-up.
|
| |
|
Event or Problem Description
|
|
During a surgical procedure, the clinician experienced an electrical pulse feedback through the renew electrocautery probe.There was no harm to the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned, decontaminated and visually inspected.There were no signs of physical damage, wear and tear to the device.A new tip was attached to the device.The device was then tested for functionality and for insulation.The device passed both tests without a defect or malfunction.No other defects were noted.This complaint is unconfirmed.
|
| |
|
Event or Problem Description
|
|
During a surgical procedure, the clinician experienced an electrical pulse feedback through the renew electrocautery probe.There was no harm to the patient or user.
|
| |
|