| Catalog Number |
SP302 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the sheaths were very sticky and hard to remove.
|
| |
|
Event or Problem Description
|
|
It was reported that the sheaths were very sticky and hard to remove.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation of the sample noted no obvious defects.An adhesive peel test was performed, the samples were cut (0.7055", 0.6905") to the required width (0.70" +/- 0.05 ").Adhesive peel strength (1.0605 lbf, 0.9542 lbf) was found to be within specification (0.60-2.10lbf).Although the reported event was unconfirmed, a potential failure mode could be ¿too high viscosity¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "fitting the sheath: the sheath is a strapless penile sheath: it has an adhesive coating inside to ensure a safe and simple procedure for fitting.If necessary trim pubic hair clean and dry pubic area.Remove sheath from its wrapping.Place rolled end over the end of the penis, leaving a small space between the end of the penis and the cup of the sheath if uncircumcised, do not retract the foreskin but allow it to remain over the glans.Avoid contact between the adhesive and the glans.Slowly unroll the sheath along the shaft of the penis, then gently squeeze the shad around the penis to ensure even adhesion.Try to avoid leaving a rolled 'collar' of sheath around the base of the penis.Finally, connect the urine collection bag.Always check that the connections are secure before use.Wear time will vary from user to user.It is recommended that a sheath b changed every 24 hours for reasons of hygiene.To remove : the sheath may be removed by gently tolling it off the penis.Avoid simply lulling the sheath off.Removing the sheath whilst having a bath or shower may increase comfort.Do not use on irritated or compromised skin.Discontinue use if irritation occurs.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.Visual evaluation of the sample noted no obvious defects.An adhesive peel test was performed, the samples were cut (0.7055", 0.6905") to the required width (1.00" +/- 0.05 ").Adhesive peel strength (1.0605 lbf, 0.9542 lbf) was found to be within specification (0.60-2.10lbf).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "fitting the sheath: the sheath is a strapless penile sheath: it has an adhesive coating inside to ensure a safe and simple procedure for fitting.If necessary trim pubic hair clean and dry pubic area remove sheath from its wrapping place rolled end over the end of the penis, leaving a small space between the end of the penis and the cup of the sheath if uncircumcised, do not retract the foreskin but allow it to remain over the glans.Avoid contact between the adhesive and the glans.Slowly unroll the sheath along the shaft of the penis, then gently squeeze the shad around the penis to ensure even adhesion.Try to avoid leaving a rolled 'collar' of sheath around the base of the penis.Finally, connect the urine collection bag.Always check that the connections are secure before use.Wear time will vary from user to user.It is recommended that a sheath b changed every 24 hours for reasons of hygiene.To remove : the sheath may be removed by gently tolling it off the penis.Avoid simply lulling the sheath off.Removing the sheath whilst having a bath or shower may increase comfort.Do not use on irritated or compromised skin.Discontinue use if irritation occurs.".
|
| |
|
Event or Problem Description
|
|
It was reported that the sheaths were very sticky and hard to remove.
|
| |
|