MAUDE Adverse Event Report: BARD ACCESS SYSTEMS LEONARD 10 FR DUAL-LUMEN CV CATHETER, PEEL-APART INTRODUCER KIT WITH SURECUFF; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS LEONARD 10 FR DUAL-LUMEN CV CATHETER, PEEL-APART INTRODUCER KIT WITH SURECUFF; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 0600630
Medical Device Problem Codes Burst Container or Vessel (1074); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that when the nurse flushed the line it ruptured.The line was replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a dhr review is not required, however it was requested.The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device returned was one 10 fr d/l 90 cm leonard catheter with surecuff complete.Functional, microscopic, tactual, and visual evaluation was performed.A w-shaped split with granular fracture surface was found immediately distal to the bifurcation, which is consistent with a burst.Therefore, the investigation is confirmed for an external leak, more specifically an external burst.The definitive root cause could not be determined based upon available information.It is unknown if patient and/or procedural issues contributed to the reported event.However, overpressurization, including use of a syringe smaller than 10 ml, is a known cause of bursts and may have contributed to this event.Labeling review: a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, cautions, possible complications and contraindications showed that the product labeling is adequate.(expiry date: 10/2022).
 
Event or Problem Description
It was reported that when the nurse flushed the line it ruptured.The line was replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEONARD 10 FR DUAL-LUMEN CV CATHETER, PEEL-APART INTRODUCER KIT WITH SURECUFF
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8033497
Report Number3006260740-2018-03084
Device Sequence Number1303275
Product Code LJS
UDI-Device Identifier00801741051944
UDI-Public(01)00801741051944
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600630
Device Catalogue Number0600630
Device Lot NumberHUCR1207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Initial Date Received by Manufacturer 10/11/2018
Supplement Date Received by Manufacturer03/18/2019
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-