• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED SENSOR MMT-7020A ENLITE3 5PK US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020A
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Inspected 1 opened/used sensor and found sensor cannula contact pad damaged (wrinkled) unable to confirm customer received sensor in said condition due to customer returned opened/used.
 
Event or Problem Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 108 mg/dl and the sensor glucose was 48 mg/dl at the time of the incident.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020A ENLITE3 5PK US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8033737
Report Number2032227-2018-55016
Device Sequence Number1643130
Product Code OZP
UDI-Device Identifier20643169704057
UDI-Public(01)20643169704057(017)20180921
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/21/2018
Device Model NumberMMT-7020A
Device Catalogue NumberMMT-7020A
Device Lot NumberI217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2018
Initial Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
Patient Weight198
-
-