| Model Number |
75974 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sample was not returned.No sample was received for evaluation but a picture was provided confirming the reported issue.Device history record (dhr) was reviewed and no issues were noted related to this complaint.The certificate of conformance (coc) was reviewed for this component and the component was deemed conforming by the supplier surge cardiovascular.100 percent pull testing was performed by the supplier to a specification of 5 lbs and all product passed this specification, there were no failures reported by the supplier.Additionally there were no issues noted in the review of the incoming inspection records.This complaint is being reviewed internally and with the supplier of the product surge cardiovascular.The results of this investigation will be updated in the follow-up mdr.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.The case label gtin: (b)(4).The product identifier: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that during a cardiopulmonary bypass (cpb) procedure, the tip of the sucker came off in the patient's chest.It happened 1 to 1.5 hours into the case and was noticed right away when it fell off.The tip was removed and the product was changed out.The surgery was completed successfully.There was no patient injury or blood loss.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up is submitted to the fda in accord with applicable regulations - and as indicated by terumo cardiovascular systems corporation in the initial report submitted november 12, 2018.(b)(4).The supplier conducted a root cause analysis and found the root cause was due to insufficient glue.Our supplier has since automated the glue dispensing process and validated this new process which controls time, pressure and vacuum of the glue application.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
|
| |
|
Search Alerts/Recalls
|