Model Number 09-1463-0120
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/16/2018
Type of Reportable Event Malfunction
Event or Problem Description
When the amplifier was powered on smoke was emitted.There were no patient or user adverse consequences.
 
Additional Manufacturer Narrative
One ep-workmate¿ amplifier was received for evaluation.As received, the outer casing of the returned amplifier had severe damage.Due to the severity of the physical damage, no further testing was performed.The reported complaint that the scu emitted smoke when powered on could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 120 CHANNEL
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key8033794
Report Number2184149-2018-00178
Device Sequence Number11121803
Product Code DQK
UDI-Device Identifier05415067003119
UDI-Public05415067003119
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number09-1463-0120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer12/04/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-