Model Number 50000000E |
Device Problems
Display or Visual Feedback Problem (1184); Inaccurate Flow Rate (1249)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the device was receiving a low flow alert.Per troubleshooting, the complaint disconnected and reconnected the pads with no change to the flow rate.She took off the fluid delivery line and ran diagnostics with no change to the flow rate.The nurse switched the patient to a second device and the flow rate improved.The first device was sent to the biomed.
|
|
Event Description
|
It was reported that the device was receiving a low flow alert.Per troubleshooting, the complaint disconnected and reconnected the pads with no change to the flow rate.She took off the fluid delivery line and ran diagnostics with no change to the flow rate.The nurse switched the patient to a second device and the flow rate improved.The first device was sent to the biomed.
|
|
Manufacturer Narrative
|
The reported issue was inconclusive.The root cause of the reported issue could not be determined.A potential root cause of the reported issue could be a defective circulation pump.However, this could not be confirmed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "the arctic sun® temperature management system is a thermal regulating system, indicated for monitoring and controlling patient temperature in adult and pediatric patients of all ages.Warnings and cautions: warnings: do not use the arctic sun® temperature management system in the presence of flammable agents because an explosion and/or fire may result.Do not use high frequency surgical instruments or endocardial catheters while the arctic sun® temperature management system is in use.There is a risk of electrical shock and hazardous moving parts.There are no user serviceable parts inside.Do not remove covers.Refer servicing to qualified personnel.Power cord has a hospital grade plug.Grounding reliability can only be achieved when connected to an equivalent receptacle marked ¿hospital use¿ or ¿hospital grade¿.When using the arctic sun® temperature management system, note that all other thermal conductive systems, such as water blankets and water gels, in use while warming or cooling with the arctic sun® temperature management system may actually alter or interfere with patient temperature control.Do not place arcticgel¿ pads over transdermal medication patches as warming can increase drug delivery, resulting in possible harm to the patient.The arctic sun® temperature management system is not intended for use in the operating room environment.Cautions: this product is to be used by or under the supervision of trained, qualified medical personnel.Federal law (usa) restricts this device to sale, by or on the order of a physician.Use only sterile water.The use of other fluids will damage the arctic sun® temperature management system.When moving the arctic sun® temperature management system always use the handle to lift the controller over an obstacle to avoid over balancing.The patient¿s bed surface should be located between 30 and 60 inches (75 cm and 150 cm) above the floor to ensure proper flow and minimize risk of leaks.The clinician is responsible to determine the appropriateness of custom parameters.When the system is powered off, all changes to parameters will revert to the default unless the new settings have been saved as new defaults in the advanced setup screen.For small patients (=30 kg) it is recommended to use the following settings: water temperature high limit =40°c (104°f); water temperature low limit =10°c (50°f); control strategy = 2.It is recommended to use the patient temperature high and patient temperature low alert settings.The operator must continuously monitor patient temperature when using manual control and adjust the temperature of the water flowing through the pads accordingly.Patient temperature will not be controlled by the arctic sun® temperature management system in manual control.Due to the system¿s high efficiency, manual control is not recommended for long duration use.The operator is advised to use the automatic therapy modes (e.G.Control patient, cool patient, rewarm patient) for automatic patient temperature monitoring and control.The arctic sun® temperature management system will monitor and control patient core temperature based on the temperature probe attached to the system.The clinician is responsible for correctly placing the temperature probe and verifying the accuracy and placement of the patient probe at the start of the procedure.Medivance supplies temperature simulators (fixed value resistors) for testing, training and demonstration purposes.Never use this device, or other method, to circumvent the normal patient temperature feedback control when the system is connected to the patient.Doing so exposes the patient to the hazards associated with severe hypo- or hyper-thermia.Medivance recommends measuring patient temperature from a second site to verify patient temperature.Medivance recommends the use of a second patient temperature probe connected to the arctic sun® temperature management system temperature 2 input as it provides continuous monitoring and safety alarm features.Alternatively, patient temperature may be verified periodically with independent instrumentation.The displayed temperature graph is for general information purposes only and is not intended to replace standard medical record documentation for use in therapy decisions.Patient temperature will not be controlled and alarms are not enabled in stop mode.Patient temperature may increase or decrease with the arctic sun® temperature management system in stop mode.Carefully observe the system for air leaks before and during use.If the pads fail to prime or a significant continuous air leak is observed in the pad return line, check connections.If needed, replace the leaking pad.Leakage may result in lower flow rates and potentially decrease the performance of the system.The arctic sun® temperature management system is for use only with the arcticgel¿ pads.The arcticgel¿ pads are only for use with the arctic sun® temperature management systems.The arcticgel¿ pads are non-sterile for single patient use.Do not reprocess or sterilize.If used in a sterile environment, pads should be placed according to the physician¿s request, either prior to the sterile preparation or sterile draping.Arcticgel¿ pads should not be placed on a sterile field.Use pads immediately after opening.Do not store pads once the kit has been opened.Do not place arcticgel¿ pads on skin that has signs of ulceration, burns, hives, or rash.While there are no known allergies to hydrogel materials, caution should be exercised with any patient who has a history of skin allergies or sensitivities.Do not allow circulating water to contaminate the sterile field when patient lines are disconnected.The water content of the hydrogel affects the pad¿s adhesion to the skin and conductivity, and therefore, the efficiency of controlling patient temperature.Periodically check that pads remain moist and adherent.Replace pads when the hydrogel no longer uniformly adheres to the skin.Replacing pads at least every 5 days is recommended.Do not puncture the arcticgel¿ pads with sharp objects.Punctures will result in air entering the fluid pathway and may reduce performance.If accessible, examine the patient¿s skin under the arcticgel¿ pads often, especially those at higher risk of skin injury.Skin injury may occur as a cumulative result of pressure, time and temperature.Possible skin injuries include bruising, tearing, skin ulcerations, blistering, and necrosis.Do not place bean bag or other firm positioning devices under the arcticgel¿ pads.Do not place positioning devices under the pad manifolds or patient lines.The rate of temperature change and potentially the final achievable patient temperature is affected by many factors.Treatment application, monitoring and results are the responsibility of the attending physician.If the patient does not reach target temperature in a reasonable time or the patient is not able to be maintained at the target temperature, the skin may be exposed to low or high water temperatures for an extended period of time which may increase the risk for skin injury.Ensure that pad sizing/ coverage and custom parameter settings are correct for the patient and treatment goals, water flow is greater than or equal to 2.3 liters per minute and the patient temperature probe is in the correct place.For patient cooling, ensure environmental factors such as excessively hot rooms, heat lamps, and heated nebulizers are eliminated and patient shivering is controlled.Otherwise, consider increasing minimum water temperature, modifying target temperature to an attainable setting or discontinuing treatment.For patient warming, consider decreasing maximum water temperature, modifying target temperature to an attainable setting or discontinuing treatment.Due to underlying medical or physiological conditions, some patients are more susceptible to skin damage from pressure and heat or cold.Patients at risk include those with poor tissue perfusion or poor skin integrity due to diabetes, peripheral vascular disease, poor nutritional status, steroid use or high dose vasopressor therapy.If warranted, use pressure relieving or pressure reducing devices under the patient to protect from skin injury.Do not allow urine, antibacterial solutions or other agents to pool underneath the arcticgel¿ pads.Urine and antibacterial agents can absorb into the pad hydrogel and cause chemical injury and loss of pad adhesion.Replace pads immediately if these fluids come into contact with the hydrogel.Do not place arcticgel¿ pads over an electrosurgical grounding pad.The combination of heat sources may result in skin burns.If needed, place defibrillation pads between the arcticgel¿ pads and the patient¿s skin.Carefully remove arcticgel¿ pads from the patient¿s skin at the completion of use.Discard used arcticgel¿ pads in accordance with hospital procedures for medical waste.The usb data port is to be used only with a standalone usb flash drive.Do not connect to another mains powered device during patient treatment.Users should not use cleaning or decontamination methods different from those recommended by the manufacturer without first checking with the manufacturer that the proposed methods will not damage the equipment.Do not use bleach (sodium hypochlorite) as it may damage the system.Medivance will not be responsible for patient safety or equipment performance if the procedures to operate, maintain, modify or service the medivance arctic sun® temperature management system are other than those specified by medivance.Anyone performing the procedures must be appropriately trained and qualified." section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that the device was receiving a low flow alert.Per troubleshooting, the complaint disconnected and reconnected the pads with no change to the flow rate.She took off the fluid delivery line and ran diagnostics with no change to the flow rate.The nurse switched the patient to a second device and the flow rate improved.The first device was sent to the biomed.Per additional information received via ms&s data with biomed mike on 10dec2018, the biomed troubleshot the arctic sun device with ms&s assistance.With the fluid delivery line attached, the flow rate was 1.2l/min, ip was -9psi, and cp was at 4%.The biomed was instructed to remove the fluid delivery line and feel for suction.He stated that he could feel suction.The biomed was advised to inspect the fluid delivery line thoroughly, and he stated that he saw no damage.He reconnected the fluid delivery line and the flow rate stabilized at 1.2l/min with ip at -7.1psi and cp at 49%.Pump hours were noted at 400.1 and system hours at 447.2.
|
|
Manufacturer Narrative
|
The reported issue was confirmed.The root cause of the reported issue was a defective circulation pump.The united was tested with a calibration assessment unit (ctu) and the unit ran over night to see if any error code occurred.The unit received an error code 61 and after resetting the code, an error 48 occurred and while attempting to clear with ctu, the unit could not calibrate due to low flow.The unit was tested with the test processor card and there was no error code 48.A test eprom was placed which did not clear up the error code 48 either.The circulation pump was replaced.After troubleshooting with a test circulation pump, the error code 48 still occurred signifying that it was the processor card that needs to be replaced.The unit was delayed due to bad part ordered out of stock.Processor card was replaced, as well as acc and power supply cards.The wire harness for the lower bracket was replaced due to a bad ground wire.The arctic sun unit passed all applicable testing after the repair.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following:
"the arctic sun® temperature management system is a thermal regulating
system, indicated for monitoring and controlling patient temperature in adult
and pediatric patients of all ages.Warnings and cautions
warnings
¿ do not use the arctic sun® temperature management system in
the presence of flammable agents because an explosion and/or fire
may result.¿ do not use high frequency surgical instruments or endocardial catheters
while the arctic sun® temperature management system is in use.¿ there is a risk of electrical shock and hazardous moving parts.There are
no user serviceable parts inside.Do not remove covers.Refer servicing to
qualified personnel.¿ power cord has a hospital grade plug.Grounding reliability can only be
achieved when connected to an equivalent receptacle marked ¿hospital
use¿ or ¿hospital grade¿.¿ when using the arctic sun® temperature management system, note that
all other thermal conductive systems, such as water blankets and water
gels, in use while warming or cooling with the arctic sun® temperature
management system may actually alter or interfere with patient
temperature control.¿ do not place arcticgel¿ pads over transdermal medication patches
as warming can increase drug delivery, resulting in possible harm to
the patient.¿ the arctic sun® temperature management system is not intended for
use in the operating room environment.Cautions
¿ this product is to be used by or under the supervision of trained, qualified
medical personnel.¿ federal law (usa) restricts this device to sale, by or on the order of
a physician.¿ use only sterile water.The use of other fluids will damage the arctic sun®
temperature management system.¿ when moving the arctic sun® temperature management system always
use the handle to lift the controller over an obstacle to avoid over balancing.¿ the patient¿s bed surface should be located between 30 and 60 inches
(75 cm and 150 cm) above the floor to ensure proper flow and minmi ize
risk of leaks.¿ the clinician is responsible to determine the appropriateness of custom
parameters.When the system is powered off, all changes to parameters will
revert to the default unless the new settings have been saved as new defaults
in the advanced setup screen.For small patients (=30 kg) it is recommended
to use the following settings: water temperature high limit =40°c (104°f);
water temperature low limit =10°c (50°f); control strategy = 2.It is
recommended to use the patient temperature high and patient temperature
low alert settings.¿ the operator must continuously monitor patient temperature when using
manual control and adjust the temperature of the water flowing through
the pads accordingly.Patient temperature will not be controlled by the
arctic sun® temperature management system in manual control.¿ due to the system¿s high efficiency, manual control is not recommended
for long duration use.The operator is advised to use the automatic therapy
modes (e.G.Control patient, cool patient, rewarm patient) for automatic
patient temperature monitoring and control.¿ the arctic sun® temperature management system will monitor and control
patient core temperature based on the temperature probe attached to the
system.The clinician is responsible for correctly placing the temperature
probe and verifying the accuracy and placement of the patient probe at the
start of the procedure.¿ medivance supplies temperature simulators (fixed value resistors) for
testing, training and demonstration purposes.Never use this device, or other
method, to circumvent the normal patient temperature feedback control
when the system is connected to the patient.Doing so exposes the patient
to the hazards associated with severe hypo- or hyper-thermia.¿ medivance recommends measuring patient temperature from a second
site to verify patient temperature.Medivance recommends the use of a
second patient temperature probe connected to the arctic sun® temperature
management system temperature 2 input as it provides continuous
monitoring and safety alarm features.Alternatively, patient temperature may
be verified periodically with independent instrumentation.¿ the displayed temperature graph is for general information purposes only
and is not intended to replace standard medical record documentation for
use in therapy decisions.¿ patient temperature will not be controlled and alarms are not enabled in
stop mode.Patient temperature may increase or decrease with the
arctic sun® temperature management system in stop mode.¿ carefully observe the system for air leaks before and during use.If the
pads fail to prime or a significant continuous air leak is observed in the
pad return line, check connections.If needed, replace the leaking pad.Leakage may result in lower flow rates and potentially decrease the
performance of the system.¿ the arctic sun® temperature management system is for use only with
the arcticgel¿ pads.¿ the arcticgel¿ pads are only for use with the arctic sun® temperature
management systems.¿ the arcticgel¿ pads are non-sterile for single patient use.Do not
reprocess or sterilize.If used in a sterile environment, pads should be
placed according to the physician¿s request, either prior to the sterile
preparation or sterile draping.Arcticgel¿ pads should not be placed on
a sterile field.¿ use pads immediately after opening.Do not store pads once the kit has
been opened.¿ do not place arcticgel¿ pads on skin that has signs of ulceration, burns,
hives, or rash.¿ while there are no known allergies to hydrogel materials, caution
should be exercised with any patient who has a history of skin allergies
or sensitivities.¿ do not allow circulating water to contaminate the sterile field when patient
lines are disconnected.¿ the water content of the hydrogel affects the pad¿s adhesion to the
skin and conductivity, and therefore, the efficiency of controlling patient
temperature.Periodically check that pads remain moist and adherent.Replace pads when the hydrogel no longer uniformly adheres to the skin.Replacing pads at least every 5 days is recommended.¿ do not puncture the arcticgel¿ pads with sharp objects.Punctures will
result in air entering the fluid pathway and may reduce performance.¿ if accessible, examine the patient¿s skin under the arcticgel¿ pads often,
especially those at higher risk of skin injury.Skin injury may occur as a
cumulative result of pressure, time and temperature.Possible skin injuries
include bruising, tearing, skin ulcerations, blistering, and necrosis.Do not
place bean bag or other firm positioning devices under the arcticgel¿ pads.Do not place positioning devices under the pad manifolds or patient lines.¿ the rate of temperature change and potentially the final achievable
patient temperature is affected by many factors.Treatment application,
monitoring and results are the responsibility of the attending physician.If the patient does not reach target temperature in a reasonable time or
the patient is not able to be maintained at the target temperature, the skin
may be exposed to low or high water temperatures for an extended period
of time which may increase the risk for skin injury.Ensure that pad sizing/
coverage and custom parameter settings are correct for the patient and
treatment goals, water flow is greater than or equal to 2.3 liters per minute
and the patient temperature probe is in the correct place.For patient
cooling, ensure environmental factors such as excessively hot rooms,
heat lamps, and heated nebulizers are eliminated and patient shivering
is controlled.Otherwise, consider increasing minimum water temperature,
modifying target temperature to an attainable setting or discontinuing
treatment.For patient warming, consider decreasing maximum water
temperature, modifying target temperature to an attainable setting or
discontinuing treatment.¿ due to underlying medical or physiological conditions, some patients are
more susceptible to skin damage from pressure and heat or cold.Patients
at risk include those with poor tissue perfusion or poor skin integrity due to
diabetes, peripheral vascular disease, poor nutritional status, steroid use
or high dose vasopressor therapy.If warranted, use pressure relieving or
pressure reducing devices under the patient to protect from skin injury.¿ do not allow urine, antibacterial solutions or other agents to pool underneath
the arcticgel¿ pads.Urine and antibacterial agents can absorb into the
pad hydrogel and cause chemical injury and loss of pad adhesion.Replace
pads immediately if these fluids come into contact with the hydrogel.¿ do not place arcticgel¿ pads over an electrosurgical grounding pad.The combination of heat sources may result in skin burns.¿ if needed, place defibrillation pads between the arcticgel¿ pads and the
patient¿s skin.¿ carefully remove arcticgel¿ pads from the patient¿s skin at the completion
of use.Discard used arcticgel¿ pads in accordance with hospital
procedures for medical waste.¿ the usb data port is to be used only with a standalone usb flash drive.Do not connect to another mains powered device during patient treatment.¿ users should not use cleaning or decontamination methods different from
those recommended by the manufacturer without first checking with the
manufacturer that the proposed methods will not damage the equipment.Do not use bleach (sodium hypochlorite) as it may damage the system.¿ medivance will not be responsible for patient safety or equipment
performance if the procedures to operate, maintain, modify or service the
medivance arctic sun® temperature management system are other than
those specified by medivance.Anyone performing the procedures must be
appropriately trained and qualified."
correction: d10, h3, h6.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|