| Model Number |
5140-4629 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Death (1802); Perforation (2001)
|
| Date of Event |
10/08/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.The involved device was not made available for return for further investigation.Based on the event description, there is no suspicion of a device malfunction.The patient death was a result of the physician's attempt to re-position the cannula without support.The device instructions for use (ifu) contain a clear warning against re-positioning, which states "do not attempt to position or reposition the cannula without the introducer in place.Damage to the cannula and/or injury to the patient may result." this event was the result of the user failing to adhere to the instructions outlined in the ifu.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
| |
|
Event or Problem Description
|
|
After inserting the protek duo veno-venous cannula and positioning the tip through the pulmonic valve into the pulmonary artery, the device slipped back and the physician decided to re-position.The cannula was pushed forward and perforated the pulmonary artery.A surgeon was called and was in the room within 3 minutes and was unable to contain the perforation.The patient expired.
|
| |
|