• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 6.5MM CANCELLOUS BONE SCREW 16MM THREAD/85MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 6.5MM CANCELLOUS BONE SCREW 16MM THREAD/85MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 216.085
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional device product code: ktt.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on (b)(6) 2018, during sterile processing, it was found out that the two (2) hexagon screw recessed of the partially threaded had a metal shaving that was broke loose and one (1) hexagon screw had a metal shaving that was still stuck in the hexagon screw.There was no patient involvement.This report is for one (1) 6.5mm cancellous bone screw 16mm thread/85mm.This is report 3 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
It was reported on (b)(6) 2018, during sterile processing, it was found out that the two (2) hexagon screw recessed of the partially threaded with a metal shavings that was broke loose and one (1) hexagon screw had a metal shaving that was still stuck in the hexagon screw.There was no patient involvement.Flow: damage: visual (appearance not as expected).Visual inspection performed at customer quality (cq) no issues were noted with the implant, there is no evidence that threads peeled/broken.On the screw head there is a piece of metal shaving embedded inside however the shaving does not appear to be part of any of the screws that were returned.There were metal shavings that came back; however they don't belong to any of the returned implants.The received condition does not agree with the complaint description and is unconfirmed.While no definitive root cause could be determined as no issues with the implant itself were discovered.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.
 
Additional Manufacturer Narrative
The device was received, and the product evaluation is in progress.No conclusion can be drawn.Additionally, device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5MM CANCELLOUS BONE SCREW 16MM THREAD/85MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8034071
Report Number2939274-2018-54752
Device Sequence Number13827550
Product Code HWC
UDI-Device Identifier10886982155283
UDI-Public(01)10886982155283
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K061621
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number216.085
Device Catalogue Number216.085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer11/07/2018
12/10/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date11/19/2018
12/11/2018
Patient Sequence Number1
-
-