| Model Number |
216.035 |
| Medical Device Problem Codes |
Break (1069); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional device product code: ktt.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2018, during sterile processing, it was found out that the two (2) hexagon screw recessed of the partially threaded had a metal shaving that was broke loose and one (1) hexagon screw had a metal shaving that was still stuck in the hexagon screw.There was no patient involvement.This report is for one (1) 6.5mm cancellous bone screw 16mm thread/35mm.This is report 1 of 3 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received, and the product evaluation is in progress.No conclusion can be drawn.Additionally, device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary complaint summary: it was reported on (b)(6) 2018, during sterile processing, it was found out that the two (2) hexagon screw recessed of the partially threaded with a metal shavings that was broke loose and one (1) hexagon screw had a metal shaving that was still stuck in the hexagon screw.There was no patient involvement.Flow: damage: visual (appearance not as expected): visual inspection: visual inspection performed at customer quality (cq) no issues were noted with the implant, there is no evidence that threads peeled/broken.On the screw head there is a piece of metal shaving embedded inside however the shaving does not appear to be part of any of the screws that were returned.There were metal shavings that came back; however they don't belong to any of the returned implants.The received condition does not agree with the complaint description and is unconfirmed.Investigation conclusion: while no definitive root cause could be determined as no issues with the implant itself were discovered.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.
|
| |
|
Search Alerts/Recalls
|