• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / XCELA; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / XCELA; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Catalog Number H965457310
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It has been reported that a sample will be returned to angiodynamics for evaluation, but to date, nothing has been received.Upon completion of the investigation, a supplemental medwatch will be submitted.((b)(4)).
 
Event or Problem Description
As reported valved xcela picc placed on october 7.Unable to draw blood during insertion.Replaced on october 8 with another picc.No patient complications were reported.
 
Additional Manufacturer Narrative
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The december 2018 angiodynamics complaint report was reviewed for the xcela pasv picc product family and the failure mode "catheter occluded {treatment}." no adverse trend was identified.The used device from the reported event was not returned to angiodynamics for evaluation, therefore the failure was unable to be confirmed and the root cause unable to be determined.Previously, however, 13 unused devices from the same product lot had been returned by the end user hospital and had been tested : the valves and catheter tubing were found to be undamaged, with the valves slit and centered.When infused with pressurized water, no leakage occurred.The lumens were subjected to a guidewire insertion test and found to be unobstructed.The devices functioned as intended.The directions for use (dfu) provided with the xcela pasv picc contains the following directions and precautions: "flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.- flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: never forcibly flush an obstructed lumen.If either lumen develops a thrombus, first attempt to aspirate the clot with a syringe.If aspiration fails, consult institutional protocol for management of thrombosis." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / XCELA
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key8034259
Report Number1317056-2018-00191
Device Sequence Number19147989
Product Code LJS
UDI-Device IdentifierH965457310
UDI-PublicH965457310
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K091261
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Catalogue NumberH965457310
Device Lot Number5380749
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer10/09/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-