| Catalog Number |
112210 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Femoral array flickered in and out throughout case; vizadiscs were swapped multiple times; burr would not start.Once array was swapped, case continued.Surgical delay: 16-30 minutes.Case type: pka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: it was reported that femoral array flickered in and out throughout case; vizadiscs were swapped multiple times; burr would not start.Once array was swapped, case continued.Product evaluation and results: product inspection could not be performed as the product was lost and not available for evaluation.Product history review: review of the device history records indicate 25 devices were manufactured and accepted into final stock on 03/29/11.No non-conformances were identified during inspection.Complaint history review: a review of complaints in trackwise related to p/n 112210, lot number 06130211 shows no additional complaint related to the failure in this investigation.Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.Device not returned.
|
| |
|
Event or Problem Description
|
|
Femoral array flickered in and out throughout case; vizadiscs were swapped multiple times; burr would not start.Once array was swapped, case continued.Surgical delay 16-30 minutes.Case type: pka.
|
| |
|