• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PER-Q-CATH PLUS PICC 4F SINGLE-LUMEN PICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PER-Q-CATH PLUS PICC 4F SINGLE-LUMEN PICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
During insertion of perq basic kit in patients the catheter along with stylet bend inside the vein and unable to do the procedure.It was stated this occurred with four devices.This report addresses the second device.
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a bend in the catheter was confirmed, and the cause appeared to be associated with use of the device.One 4 fr s/l per-q-cath picc was returned for investigation.The catheter was received with the stylet in the lumen.The stylet had been retracted through the catheter.The distal tip of the stylet coincided with the 49cm depth mark.The catheter length had not been modified and extended to the 65cm depth mark.It is unknown if an attempt was made to insert the catheter with the stylet retracted in the lumen.The stylet was bent just proximal to the septum on the t-lock extension set.Another bend in the stylet caused a bend in the catheter near the 12cm depth mark.A functional test revealed no leaks in the catheter.The stylet was within specification.It was reported that the catheter and stylet bent during insertion.Due to the condition of the catheter, it appears that the damage occurred during use and could be associated with the insertion technique, insertion angle, or patient position.
 
Event or Problem Description
During insertion of perq basic kit in patients the catheter along with stylet bend inside the vein and unable to do the procedure.It was stated this occurred with four devices.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PER-Q-CATH PLUS PICC 4F SINGLE-LUMEN PICC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8034642
Report Number3006260740-2018-03119
Device Sequence Number19207261
Product Code LJS
UDI-Device Identifier00801741074875
UDI-Public(01)00801741074875
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K954104
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number4134105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date01/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-