Model Number VL2610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 06/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the procedure, the patient had burns in various parts of the body specifically at the arms, legs and foot.Apparently, there was leakage of energy and the burns occurred due to the inadequate substance used for asepsis with alcoholic content.It was identified that there was an automatic activation when the pedal was connected to the equipment, as there was poor contact in the pedal cable.
 
Additional Manufacturer Narrative
Additional information:evaluation summary: the product was not returned; however, a picture was provided by the customer for analysis.The product was not included in the pictures provided.The picture showed burns on a patient.Without the product, a detailed investigation could not be performed.The file will be closed as unconfirmed at this time.If additional information is obtained, or the product is returned, we will re-open this investigation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow drive
boulder CO 80301
MDR Report Key8034716
Report Number1717344-2018-01432
Device Sequence Number1435413
Product Code GEI
UDI-Device Identifier10884521196636
UDI-Public10884521196636
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K982743
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model NumberVL2610
Device Catalogue NumberVL2610
Device Lot Number1611095X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2018
Supplement Date Received by Manufacturer11/30/2018
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
-
-