| Catalog Number |
06L45-41 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Mdr number 1628664-2018-01983 has been submitted for this issue that also involves the architect c8000 analyzer.An evaluation is in process.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely depressed results for a pediatric patient while using clinical chemistry total bilirubin reagents with the architect c8000 analyzer.The following data was provided.The customer uses normal range less than or equal to 1.9 ng/ml.Sid (b)(4), initial 0.7 ng/dl, repeat 6.8, 6.8 ng/dl.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the suspect medical device was determined to not be the reagent.The suspect medical device was determined to be only the architect c8000 analyzer.Mdr number 1628664-2018-01983 was previously submitted and all further information will be documented under that mdr number.
|
| |
|