Model Number Z9002
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown, not provided.Implant and explant dates: if implanted or explanted, give date: not applicable, as lens was not implanted.All pertinent information available to johnson and johnson vision has been submitted.
 
Event or Problem Description
It was reported that the tecnis monofocal intraocular lens (model z9002 +16.5 diopter) had a bent haptic.There was no patient involvement.No further information provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 10/26/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The lens was received inside the lens case in the packaging box.The lens was observed under microscope and with viscoelastic residue in the optic zone.One of the haptics was observed positioned and the other haptic was observed distorted with the drill hole slightly torn.This condition could be caused by the handpiece pushrod.The complaint was verified in the sample returned; however, it could be related to the handling process.A product quality deficiency was not identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8035384
Report Number2648035-2018-01430
Device Sequence Number13819382
Product Code HQL
UDI-Device Identifier05050474530621
UDI-Public(01)05050474530621(17)200922
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2020
Device Model NumberZ9002
Device Catalogue NumberZ900200165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/05/2018
Supplement Date Received by Manufacturer11/12/2018
10/25/2020
Initial Report FDA Received Date11/02/2018
Supplement Report FDA Received Date11/20/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-