• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number 1012532-100
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Visual inspection was performed on the returned device.The reported complaint was confirmed.Fourier transform infrared spectrometer (ftir) testing was done on the loose material.The investigation determined the loose material in the sealed header bag appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that during incoming inspection of an 10x100mm absolute pro vascular self-expanding stent system (sess) a small piece of white foreign material was found inside the sterile pouch.The inner pouch was intact and the sterility of the device was not compromised.The incident occurred during an inspection of a consignment return at the distribution center.The sess was not used.There was no patient involvement.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8035432
Report Number2024168-2018-08439
Device Sequence Number1464049
Product Code NIO
UDI-Device Identifier08717648175879
UDI-Public08717648175879
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number1012532-100
Device Lot Number6082261
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2018
Initial Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-