| Model Number |
M00584010 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a dreamtome rx 49 was used for a sphincterotomy on (b)(6) 2018.According to the complainant, during the procedure, it was noticed that after easy catheterization, the cutting wire broke and came out of its place.There were no reported patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a dreamtome rx 49 was used for a sphincterotomy on october 4, 2018.According to the complainant, during the procedure, it was noticed that after easy catheterization, the cutting wire broke and came out of its place.There were no reported patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: problem code 1069 captures the reportable event of wire broke.Block h10: visual evaluation of the returned device found the unit returned has the cutting wire broken and blackened, consequently, confirming the reported complaint.According to the product analysis, the cutting wire of the returned product was found to be broken and blackened, so it is most likely that a peak of voltage could have caused the failures noted.Therefore, it was concluded that the investigation conclusion code of this event is "cause traced to component failure" since the failure was an expected or random component failure without any design or manufacturing issue.A search of the complaint database revealed that no other complaints exist for the specified lot.-¿the device history record review found the device met all manufacturing specifications.This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 12/19/2018 11:15 pm ct.The report number is being corrected from: 3005099803-2018-61978 to: 3005099803-2018-61062.
|
| |
|