• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX; DQY Back to Search Results
Catalog Number PNML6F0881004
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the neuron max was ovalized approximately 56.0, 90.0, and 101.5 thru 102.5 cm from the hub.Resistance was encountered while advancing the returned 6f select through the returned neuron max and the 6f select could not be advanced any further.Conclusions: evaluation of the returned neuron max revealed that the device was ovalized.If the device is forcefully gripped or pinched during preparation, damage such as this may occur.The ovalization in the neuron max likely contributed to the 6f select¿s inability to advance through the sheath during preparation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.This report is associated with mfr report number: 3005168196-2018-02169.
 
Event or Problem Description
During preparation for a medical procedure, a neuron 6f select catheter (6f select) would not advance all the way through the distal tip of a neuron max 6f 088 long sheath (neuron max) on the back table; therefore, both neuron max and 6f select were not used in the procedure.The procedure was completed using a new neuron max and 6f select.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8035661
Report Number3005168196-2018-02170
Device Sequence Number19158142
Product Code DQY
UDI-Device Identifier00814548012292
UDI-Public00814548012292
Combination Product (Y/N)Y
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/03/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/12/2020
Device Catalogue NumberPNML6F0881004
Device Lot NumberF79494
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2018
Initial Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-