Results: the canister was undamaged.The demonstration pump max had a vacuum of approximately -29.0 inhg while powered on with a pressure gauge attached.When attaching the canister, the demonstration pump max still had a vacuum of approximately -29.0 inhg.The pressure gauge, however, only had a vacuum of -27.5 inhg, which is within the specification for the canister.Conclusion: evaluation of the returned canister revealed an undamaged device.When functionally tested with a demonstration pump max, the canister was able to hold a vacuum pressure within specification.Penumbra canisters are inspected during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
The patient was undergoing a thrombectomy procedure on the posterior tibial artery using an indigo pump max canister (canister).During the procedure, a penumbra staff noticed the vacuum produced by the penumbra system aspiration pump max (pump max) went down to -29inhg.After troubleshooting, the penumbra staff decided to replace the canister with a new one and full vacuum power was achieved.There was no report of an adverse effect to the patient.
|