| Model Number |
FPRPR3560 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned and could not be investigated.The manufacturing records were reviewed and no related deviation or concerns were found.Vacuum sleeve failure is a known potential malfunction with some cryoablation needles.The risk to the patient is the potential for a skin burnt due to frost on the needle shaft.Galil medical's labeling includes precautions instructing users to protect the skin with warm saline, when appropriate.The case was completed with no patient injury.If additional information becomes available a followup report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the doctor noticed the shaft of one of the icesphere needle started to collect frost.The procedure was completed as expected with no injury to the patient.
|
| |
|