Catalog Number 9392507
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with herniated intervertebral disc and underwent spinal decompression surgery.During surgery, the spacer was broken.The product came in contact with the patient.No fragments of the broken spacer remained in the patient.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and microscopic examination of the implant displays a brittle fracture with rays emanating towards the inserter interface, and deformation at the inserter attachment hole.This type of fracture is consistent with pressure being placed on the implant during a bending moment and excessive force during implant impaction leading to the overload of the material.The angulation of the fracture lines is also constant with this type of failure.The observations suggest that there may have been insufficient distraction or preparation of the disc space prior to attempted insertion of the implant.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRESCENT SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key8036729
Report Number1030489-2018-01443
Device Sequence Number1498908
Product Code MAX
UDI-Device Identifier00613994187451
UDI-Public00613994187451
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K172199
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2025
Device Catalogue Number9392507
Device Lot NumberH5349797
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer02/04/2019
Initial Report FDA Received Date11/05/2018
Supplement Report FDA Received Date02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-