| Model Number |
MMT-523RNAP |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump received with broken battery tube thread and reservoir tube lip completely broken off noted.Unable to perform displacement test due to reservoir lip broken off.The test reservoir does not stay locked in place due to reservoir tube lip broken off.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that reservoir compartment broken.The customer¿s blood glucose level was 215 mg/dl.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|