| Model Number |
EGIATRS45AXT |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when they inserted the reload during a laparoscopic sleeve gastrectomy, the arrows would not align, and a piece at the base of the reload was loose and then dislodged, and it broke off.They used another reload to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual examination of the staple cartridge noted that the reload had a full complement of staples.The lock ring was observed to be broken.Functionally the reload was loaded into a pmv representative instrument and was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damaged reload lock ring may occur due improper orientation of the lock ring or over flexure of the reload while attached to the instrument.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|