• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DISPENSING PIN; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DISPENSING PIN; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number 413500
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as (b)(4).The device involved has been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was reported that particulate matter was seen floating in bag in use with the dispensing pin.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) unused and unopened dp3500 disp pin ls reflux with packaging was returned for further evaluation.Visual examination of the dispensing pin notes no white particulates on the dispensing pin or within the blister.The defect of white particulates/plastic shavings is not confirmed.Based on the results of the sample evaluation, no specific conclusions can be made regarding the cause of the reported event.While we are unable to determine the root cause of the reported defect, we will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPENSING PIN
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key8037422
Report Number2523676-2018-00095
Device Sequence Number16502757
Product Code LHI
UDI-Device Identifier04022495769749
UDI-Public04022495769749
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K925401
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number413500
Device Lot Number0061618354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date11/05/2018
Supplement Report FDA Received Date11/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-