COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 8886848808
Device Problems Mechanical Problem (1384); Failure to Fire (2610)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic right hemicolectomy procedure, on the second firing when the surgeon clip the vessel, the clip could not be release from the cartridge.Although the surgeon managed to move it out, clips were malformed.The surgical time was extended by less than 30 min.The procedure was completed with another device.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted the cartridge was received applied with the clip body disengaged and the tracks were not inserted properly at the distal end of the clip body.Functionally; the clip body and tracks were reloaded into the cartridge and fired with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the observed condition may occur if the clip was not properly applied and mishandled during the clinical application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8037632
MDR Text Key126157490
Report Number1219930-2018-05842
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K925602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8886848808
Device Catalogue Number8886848808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Initial Date Manufacturer Received 10/10/2018
Initial Date FDA Received11/05/2018
Supplement Dates Manufacturer Received11/27/2018
Supplement Dates FDA Received12/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-