The concerto coil was not returned for analysis; therefore, no definitive conclusion can be drawn regarding the clinical observation.However, the device is pending return.Upon receipt of the device and/or additional information a supplemental report will be filed.Related mdrs for this event: 2029214-2018-00914 2029214-2018-00915.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received information that two medtronic implant coils migrated from the target location and possibly contributed to newly developed pulmonary thrombosis.Prior to the event, the two medtronic implant coils were implanted in the intended location; however, they both fell out of the aneurysm and migrated afterwards.One implant coil was found in the vena cava and the other implant coil was found in the pulmonary artery.The physicians successfully extracted both implant coils from the patient.The complication mentioned above as noticed at the time.The patient required additional hospitalization time due to the re-intervention and administration of the anticoagulants.The reported devices and accessory devices were reported to be prepared as indicated in the instructions for use (ifu).The patient¿s vasculature was moderate in tortuosity and blood flow was normal.
|
|
The device was received with the implant coil only; inside of a biohazard bag and a shipping box.There was no pusher, microcatheter, or accessories returned with the device.The implant coil was found to be damaged and stretched with the polypropylene filament broken.The detached element was not returned and missing from the proximal end of the coil.No other anomalies were observed.Based on the customer's photo, the customer complaint was confirmed.However, the event cause could not be determined.In addition, the returned coil was found damaged and stretched with the polypropylene filaments broken.However, the cause for damage could not be determined.Since the pusher, detached element and microcatheter were not returned; any contribution of the pusher, detached element and microcatheter could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|