• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 3948L
Medical Device Problem Codes Break (1069); Expulsion (2933)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a sigmoidectomy procedure, on the first firing when the surgeon approached to the colon and tried to fire the device reload got apart, in addition the other reload pre-fired during storage.The procedure was completed with another device.There was no patient injury.
 
Additional Manufacturer Narrative
Additional info: evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The loading unit was received fully applied.The knife assembly was di sengaged.The pusher thumb piece was intact.Visual inspection noted the second single use loading unit(sulu) received fully applied with no visual abnormalities.Functional testing was precluded due to the observed damage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition occurs when the firing handle is over retracted after firing.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8038575
Report Number2647580-2018-05320
Device Sequence Number1421460
Product Code GDW
UDI-Device Identifier10884521063150
UDI-Public10884521063150
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model Number3948L
Device Catalogue Number3948L
Device Lot NumberP7L1128X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer01/08/2019
Initial Report FDA Received Date11/05/2018
Supplement Report FDA Received Date01/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-