Model Number FX4308MLC
Medical Device Problem Codes Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 10/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The dental implant fx4308mlc was removed on (b)(6) 2018 due to loss of osseointegration around 9 years after implantation.The patient is a tobacco user and has history of bruxism and hypertension.The doctor noted bone resorption during explantation.The patient has recovered without complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTIUM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DENTIUMUSA
6761 katella ave
cypress CA 90630
MDR Report Key8039032
Report Number3005503242-2018-00165
Device Sequence Number1421518
Product Code DZE
UDI-Device IdentifierD755FX4308MLC0
UDI-PublicD755FX4308MLC0
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K041368
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2010
Device Model NumberFX4308MLC
Device Catalogue NumberFX4308MLC
Device Lot Number07K2301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2018
Initial Report FDA Received Date11/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
-
-