ALLERGAN (COSTA RICA) STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MX-16-T
Medical Device Problem Codes Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Cellulitis (1768)
Date of Event 08/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of cellulitis is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: deflation-patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.Use error-do not reuse explanted products.Natrelle® 133 tissue expanders are intended for single use only.
 
Event or Problem Description
Patient reported right side acute cellulitis, implant is hard and red, there may be a leak, a lot of bleeding, aspirated 300cc of blood, yellow and cloudy liquid, and the doctor "removed the expander, rinsed and replaced it".Device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MX TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8039417
Report Number9617229-2018-08497
Device Sequence Number1464282
Product Code LCJ
UDI-Device Identifier10888628004511
UDI-Public10888628004511
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/23/2022
Device Catalogue Number133MX-16-T
Device Lot Number3191802
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2018
Initial Report FDA Received Date11/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-