It was reported that patient's left humerus was revised due to a broken plate.Surgeon reported patient had suffered two falls prior to revision.The plate and 7 screws were revised to an axsos 5.0 broad compression plate.Rep reported that x-rays, medical records, and further information are not available due to hospital policy.
|
The reported event that locking compression plate axsos 3 ti 9 hole / l123mm 4.0mm locking was alleged of 'implant breakage after surgery' could be confirmed.Based on investigation, the root cause was attributed to be patient related.The failure was caused by the falls that the patient had.Indeed, the implant is not supposed to withstand forces related to important shocks.The device inspection revealed the following: the plate was returned, and is indeed broken at the level of the fifth hole.When put together, the two pieces show that the plate was severely bent, probably caused by one of the falls.The analysis of the breakage surface shows very clear breakage lines, and a large tear-off surface.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
It was reported that patient's left humerus was revised due to a broken plate.Surgeon reported patient had suffered two falls prior to revision.The plate and 7 screws were revised to an axsos 5.0 broad compression plate.Rep reported that x-rays, medical records, and further information are not available due to hospital policy.
|