| Model Number |
9733734 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ratcheting handle was returned to the manufacturer for analysis.Analysis found that the part retention mechanism had a broken retainer ring.As a result, the retention cover and spring has separated from the handle.Analysis found that the reported event was related to a mechanical issue.Device manufacturing date is unavailable.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that when the manufacturer representative received the nac navlock tray back, the ratcheting handle was damaged.There was no patient present when this issue was identified.
|
| |
|