| Model Number |
Z9002 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Implant and explant dates: if implanted or explanted, give date: not applicable, as only the haptic touch the eye in the initial surgery.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that when inserting the z9002 intraocular lens (iol) into the patient's right eye, the surgeon noticed a crack in the lens.The haptic touched the patient's eye.There was no medical or surgical intervention required.Another z9002, same diopter, was implanted as a replacement.The patient was reported doing good post-op.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 11/21/2018.Returned to manufacturer: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection showed that the product was received with the lens cut and a haptic detached.Viscoelastic residue was noted.The complaint issue reported as iol torn was confirmed; however, due the condition of the returned lens the complaint could not be related to the manufacturing process; it could be related to the handling process.A product quality deficiency was not identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Another complaint was found under this production order; however, was not related to this complaint.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|