| Model Number |
ETBF2313C145E |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Codes |
Thrombus (2101); No Consequences Or Impact To Patient (2199); Injury (2348)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: etlw1613c124e, serial/lot #: (b)(4), ubd: 09-jul-2020, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 66mm abdominal aortic aneurysm.It was reported during follow-up approximately one month later that limb occlusion was identified.Per the physician, the cause of the event is undetermined.No additional clinical sequelae were reported and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A thrombectomy was performed to clean out the thrombus and an additional stent was implanted to reline the limb and resolve the event.The patient status has been confirmed as good.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated: fdc code added at investigation completion.If information is provided in the future, a supplemental report will be issued.
|
| |
|