Model Number TRUEMETRIX
Medical Device Problem Codes Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 10/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: customer's condition has improved.Customer states that he is feeling much better- no longer experiencing symptoms.Customer went to the er on (b)(6) 2018.Customer was not hospitalized.Customer stated that when he arrived at the hospital, they took his blood sugar and a urine test.Customer was released the same day and given an rx for metformin.Customer was diagnosed with type 2 diabetes.Customer stated that they just took a urine and blood test.Results were 271mg/dl fasting.Customer was discharged the same day.No further medical attention is required.
 
Event or Problem Description
Consumer reported complaint for e-3 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.Customer not feeling physically well and report symptoms of blurry vision and dry mouth.Customer has not been diagnosed but will visit a doctor tomorrow for treatment for hyperglycemia.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room.During call on (b)(6) 2018, back to back blood test was performed non-fasting by the customer and produced test results of 341mg/dl using true metrix meter.The test strip lot manufacturer's expiration date is date and open vial date is date.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8041540
Report Number1000113657-2018-01136
Device Sequence Number1643608
Product Code NBW
UDI-Device Identifier00021292006082
UDI-Public(01)00021292006082
Combination Product (Y/N)N
Initial Reporter StateVA
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2905
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2018
Initial Report FDA Received Date11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-